What modules are available for the Sniffer4D Nano2+ multi-gas detection unit?
Up to 5 internal modules can be installed inside a Sniffer4D Nano2+ base unit. Choose the modules that fit your application.
Module | Specs | Purpose |
|---|---|---|
Inhalable Particulate Matter (PM2.5&10) Sensing Module | • Detection method: Laser scattering/light scattering. • Sense PM1.0 (particle size 0.3~1um), PM2.5 (particle size 0.3~2.5um), and PM10 (particle size 0.3~10um). • Particle counting effectiveness: 50% @ 0.3um, 98% @ > 0.5um. • Range: 0~1000ug/m3. • Detection limit: 1ug/m3. • Repeatability: <2% FS. • Theoretical resolution: 1ug/m3. • Warm-up time from a cold start: <10s. • Overall response time: <10s. • Estimated service life: >36 months. • On-chip proprietary humidity correction algorithm, enabling better data quality in wide humidity range. | For general environmental monitoring. Cannot be chosen in conjunction with TSP. |
Total Suspension Particular Matter (TSP/M100) Sensing Module | • Detection method: Laser scattering/light scattering. • Sense PM2.5 (particle size 0.3~2.5um), PM10 (particle size 0.3~10um) and PM100 (TSP) (particle size 1~100um). • Particle counting effectiveness: 50% @ 0.3um, 98% @≥0.5um. • TSP effective range: 0~2000μg/m3. • TSP maximum range: >10000μg/m3. • Theoretical resolution: 1ug/m3. • Time resolution: 1s. • On-chip proprietary humidity correction algorithm, enabling better data quality in wide humidity range. • Overall response time: ≤10s. • Estimated service life: ≥20000 h; • Operating temperature: -30 to 70°C. • Operating humidity: 0-95%RH (non-condensing). | Cannot be chosen with any other sensing modules due to limited space. |
High-resolution 03+NO2 Sensing Module | • Detection method: Electrochemistry. • Sensitive to both 03 and NO2, but unable to identify individual concentrations. • Range: 0~11ppm. • Detection limit: 5ppb. • Repeatability: <4%FS. • Overall response time (t90): <45s (0~1ppm). • Theoretical resolution: <1ppb). • On-chip proprietary humidity correction algorithm, enabling better data quality in wide humidity range. • Warm-up time from a cold start: ≈2minutes. • Sensitivity drift: -20~-40%/year (in laboratory environment). • Zero drift: 0~20ppb/year (in laboratory environment). • Est. service life: >24months. •Operating temperature: -30 to 40°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-85%RH. | For general environmental monitoring. This combination is also called Ox or photochemical oxidant, which represents the oxidising ability of the air. Individual O3 concentration is calculated using O3=(O3+NO2)-NO2. |
High-resolution NO2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~11ppm. • Detection limit: 5ppb. • Repeatability: <4%FS. • Overall response time (t90): <60s (0~2ppm). • Theoretical resolution: ~1.1ppb. • On-chip proprietary humidity correction algorithm, enabling better data quality in wide humidity range. • Warm-up time from a cold start: ~2minutes. • Sensitivity drift: -20~-40%/year (in laboratory environment). • Zero drift: 0~20ppb/year (in laboratory environment). • Est. service life: >24months. • Operating temperature: -30 to 40°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-85%RH | For general environmental monitoring, HAZMAT response, and ship fuel sulfur content monitoring. |
High-resolution CO Sensing Module | • Detection method: Electrochemistry. • Range: 0~11ppm. • Detection limit: 5ppb. • Repeatability: <4%FS. • Overall response time (t90): <20s (0~10ppm). • Theoretical resolution: ~3ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈2minutes. • Sensitivity drift: <10%/year (in laboratory environment). • Zero drift: <±100ppb/year (in laboratory environment). • Est. service life: >36months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For general environmental monitoring and HAZMAT response. |
High-resolution SO2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~15ppm. • Detection limit: 5ppb. • Repeatability: <4%FS. • Overall response time (t90): <40s (0~2ppm). • Theoretical resolution: <1ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈2minutes. • Sensitivity drift: <±15%/year (in laboratory environment). • Zero drift: <+20ppb/year (in laboratory environment). • Est. service life: >36months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90% RH. | For general environmental monitoring, HAZMAT response, and ship fuel sulfur content monitoring. |
Wide-range H2S Sensing Module | • Detection method: Electrochemistry. • Range: 0~90ppm. • Detection limit: 20ppb. • Repeatability: <4%FS. • Overall response time (t90): <55s (0~2ppm). • Theoretical resolution: ~5ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈3minutes. • Sensitivity drift: <20%/year (in laboratory environment). • Zero drift: <+100ppb/year (in laboratory environment). • Est. service life: >24months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For general environmental monitoring, oil and gas leak detection, and HAZMAT response. |
Wide-range O2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~50%. • Detection limit: 0.5%. • Overall response time (t90): <200s (20.9% -0%). • Theoretical resolution: <0.1%; • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ~60s. • Est. service life: >24months. • Operating temperature: -30 to 55°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 5-95%RH. • Operating pressure: 800~1200hPa. | For HAZMAT response |
Wide-range NH3 Sensing Module | • Detection method: Electrochemistry. • Range: 0~100ppm. • Detection limit: 1ppm. • Repeatability: <2%FS. • Accuracy: ±5%FS. • Overall response time (t90): <50s (0~50ppm). • Theoretical resolution: 0.1ppm. • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Sensitivity drift: <3%/year (in laboratory environment). • Zero drift: <+2ppm/year (in laboratory environment). Est. service life: >24months. • Operating temperature: -40 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-95%RH. | For odor detection and HAZMAT response. |
Wide-range HCl Sensing Module |