What modules are available for the Sniffer4D Nano2+ multi-gas detection unit?
Up to 5 internal modules can be installed inside a Sniffer4D Nano2+ base unit. Choose the modules that fit your application.
Module | Specs | Purpose |
|---|---|---|
Inhalable Particulate Matter (PM2.5&10) Sensing Module | • Detection method: Laser scattering/light scattering. • Sense PM1.0 (particle size 0.3~1um), PM2.5 (particle size 0.3~2.5um), and PM10 (particle size 0.3~10um). • Particle counting effectiveness: 50% @ 0.3um, 98% @ > 0.5um. • Range: 0~1000ug/m3. • Detection limit: 1ug/m3. • Repeatability: <2% FS. • Theoretical resolution: 1ug/m3. • Warm-up time from a cold start: <10s. • Overall response time: <10s. • Estimated service life: >36 months. • On-chip proprietary humidity correction algorithm, enabling better data quality in wide humidity range. | For general environmental monitoring. Cannot be chosen in conjunction with TSP. |
Total Suspension Particular Matter (TSP/M100) Sensing Module | • Detection method: Laser scattering/light scattering. • Sense PM2.5 (particle size 0.3~2.5um), PM10 (particle size 0.3~10um) and PM100 (TSP) (particle size 1~100um). • Particle counting effectiveness: 50% @ 0.3um, 98% @≥0.5um. • TSP effective range: 0~2000μg/m3. • TSP maximum range: >10000μg/m3. • Theoretical resolution: 1ug/m3. • Time resolution: 1s. • On-chip proprietary humidity correction algorithm, enabling better data quality in wide humidity range. • Overall response time: ≤10s. • Estimated service life: ≥20000 h; • Operating temperature: -30 to 70°C. • Operating humidity: 0-95%RH (non-condensing). | Cannot be chosen with any other sensing modules due to limited space. |
High-resolution 03+NO2 Sensing Module | • Detection method: Electrochemistry. • Sensitive to both 03 and NO2, but unable to identify individual concentrations. • Range: 0~11ppm. • Detection limit: 5ppb. • Repeatability: <4%FS. • Overall response time (t90): <45s (0~1ppm). • Theoretical resolution: <1ppb). • On-chip proprietary humidity correction algorithm, enabling better data quality in wide humidity range. • Warm-up time from a cold start: ≈2minutes. • Sensitivity drift: -20~-40%/year (in laboratory environment). • Zero drift: 0~20ppb/year (in laboratory environment). • Est. service life: >24months. •Operating temperature: -30 to 40°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-85%RH. | For general environmental monitoring. This combination is also called Ox or photochemical oxidant, which represents the oxidising ability of the air. Individual O3 concentration is calculated using O3=(O3+NO2)-NO2. |
High-resolution NO2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~11ppm. • Detection limit: 5ppb. • Repeatability: <4%FS. • Overall response time (t90): <60s (0~2ppm). • Theoretical resolution: ~1.1ppb. • On-chip proprietary humidity correction algorithm, enabling better data quality in wide humidity range. • Warm-up time from a cold start: ~2minutes. • Sensitivity drift: -20~-40%/year (in laboratory environment). • Zero drift: 0~20ppb/year (in laboratory environment). • Est. service life: >24months. • Operating temperature: -30 to 40°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-85%RH | For general environmental monitoring, HAZMAT response, and ship fuel sulfur content monitoring. |
High-resolution CO Sensing Module | • Detection method: Electrochemistry. • Range: 0~11ppm. • Detection limit: 5ppb. • Repeatability: <4%FS. • Overall response time (t90): <20s (0~10ppm). • Theoretical resolution: ~3ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈2minutes. • Sensitivity drift: <10%/year (in laboratory environment). • Zero drift: <±100ppb/year (in laboratory environment). • Est. service life: >36months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For general environmental monitoring and HAZMAT response. |
High-resolution SO2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~15ppm. • Detection limit: 5ppb. • Repeatability: <4%FS. • Overall response time (t90): <40s (0~2ppm). • Theoretical resolution: <1ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈2minutes. • Sensitivity drift: <±15%/year (in laboratory environment). • Zero drift: <+20ppb/year (in laboratory environment). • Est. service life: >36months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90% RH. | For general environmental monitoring, HAZMAT response, and ship fuel sulfur content monitoring. |
Wide-range H2S Sensing Module | • Detection method: Electrochemistry. • Range: 0~90ppm. • Detection limit: 20ppb. • Repeatability: <4%FS. • Overall response time (t90): <55s (0~2ppm). • Theoretical resolution: ~5ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈3minutes. • Sensitivity drift: <20%/year (in laboratory environment). • Zero drift: <+100ppb/year (in laboratory environment). • Est. service life: >24months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For general environmental monitoring, oil and gas leak detection, and HAZMAT response. |
Wide-range O2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~50%. • Detection limit: 0.5%. • Overall response time (t90): <200s (20.9% -0%). • Theoretical resolution: <0.1%; • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ~60s. • Est. service life: >24months. • Operating temperature: -30 to 55°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 5-95%RH. • Operating pressure: 800~1200hPa. | For HAZMAT response |
Wide-range NH3 Sensing Module | • Detection method: Electrochemistry. • Range: 0~100ppm. • Detection limit: 1ppm. • Repeatability: <2%FS. • Accuracy: ±5%FS. • Overall response time (t90): <50s (0~50ppm). • Theoretical resolution: 0.1ppm. • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Sensitivity drift: <3%/year (in laboratory environment). • Zero drift: <+2ppm/year (in laboratory environment). Est. service life: >24months. • Operating temperature: -40 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-95%RH. | For odor detection and HAZMAT response. |
Wide-range HCl Sensing Module | • Detection method: Electrochemistry. • Range: 0~100ppm. • Detection limit: 1ppm. • Repeatability: <4%FS. • Overall response time (t90): <200s (0~25ppm). • Theoretical resolution: <15ppb. • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈5minutes. • Est. service life: >24months. Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For oil and gas leak detection and HAZMAT response. |
Wide-range HCN Sensing Module | • Detection method: Electrochemistry. • Range: 0~100ppm. • Detection limit: 50ppb. • Repeatability: <5%FS. • Overall response time (t90): <120s (0~30ppm). • Theoretical resolution: <0.1ppm. • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ~5minutes. • Est. service life: >12months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For HAZMAT response. |
Wide-range HF Sensing Module | • Detection method: Electrochemistry. • Range: 0~100ppm. • Detection limit: 1ppm. • Repeatability: <4%FS. • Overall response time (t90): <200s (0~25ppm). • Theoretical resolution: ~20ppb. • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈5minutes. • Est. service life: >24months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | |
Wide-range H2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~5000ppm. • Detection limit: 17ppm. • Repeatability: <5%FS; • Overall response time (t90): <55s (0~400ppm). • Theoretical resolution: <0.7ppm. • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈5minutes. • Zero drift: <±20ppb/year (in laboratory environment). • Est. service life: >24months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For H2 leakage monitoring in power station accidents. |
Cl2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~20ppm. • Detection limit: 0.5ppm. • Repeatability: <4%FS. • Overall response time (t90): <60s (0~10ppm). • Theoretical resolution: ~20ppb. • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ~5minutes. • Est. service life: >24months. • Operating temperature: -20 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For HAZMAT response. |
Wide-range PH3 Sensing Module | • Detection method: Electrochemistry. • Range: 0~2000ppm. • Detection limit: 20ppm. • Repeatability: <5%FS. • Overall response time (t90): <30s (0~800ppm). • Theoretical resolution: ~0.3ppm; • On-chip proprietary individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ~5minutes. • Sensitivity drift: <4%/year (in laboratory environment). • Est. service life: >24months. • Zero drift: <1.5ppm/year (in laboratory environment). • Operating temperature: -20 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 20-90%RH. | Commonly used to check the phosphine gas emitted in the process of drug production. |
High-resolution NO Sensing module | • Detection method: Electrochemistry. • Range: 0~11ppm. • Detection limit: 5ppb. • Repeatability: <4%FS; sing Module. • Overall response time (t90): <60s (0~10ppm). • Theoreticalresolution: <1.1ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈5minutes. • Sensitivity drift: -20~-40%/year (in laboratory environment). • Zero drift: 0~20ppb/year (in laboratory environment). • Est. service life: >24 months. • Operating temperature: -30 to 40°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment.). • Operating humidity: 15-85%RH | For general environmental monitoring and HAZMAT response. |
Wide-range SO2 Sensing Module | • Detection method: Electrochemistry. • Range: 0~100ppm. • Detection limit: 50ppb. • Repeatability: <4%FS. • Overall response time (t90): <40s (0~2ppm). • Theoretical resolution: <8ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈2minutes. • Sensitivity drift: <±15%/year (in laboratory environment). • Zero drift: <+20ppb/year (in laboratory environment). • Est. service life: >36months. • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | For HAZMAT response. |
Wide-range CO Sensing Module | • Detection method: Electrochemistry. • Range: 0~1000ppm. • Detection limit: 70ppb. • Repeatability: <4%FS. • Overall response time (t90): <20s (0~10ppm). • Theoretical resolution: ~50ppb. • On-chip proprietary environmental and individual difference compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: ≈2minutes. • Sensitivity drift: <10%/year (in laboratory environment). • Est. service life: >36 months. • Zero drift: <±100ppb/year (in laboratory environment). • Operating temperature: -30 to 50°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-90%RH. | |
OU Sensing Module | • Detection method: Electrochemistry. • Range: 0~10ppm. • Detection limit: ~0.1ppm. • Repeatability: <5%FS. • Overall response time (t90): <30s (0~10ppm). • Theoretical resolution: 0.01ppm. • Warm-up time from a cold start: ~3minutes. • Est. service life: >36 months. • Operating temperature: -40°C to +55°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-95% RH (non-condensing). | Commonly used to check the phosphine OU emitted in environmental inspection. |
Wide-range CxHy/CH4/LEL Sensing Module | • Detection method: Non-dispersive infrared (NDIR). • Target gases: hydrocarbons (flammable gases). • Range: 0~5%VOL (0~100%LEL) methane, or 0~2%VOL propane. • Detection limit: 0.01%/100ppm. • Repeatability: <2%FS. • Accuracy: ±10%. • Response time (t90): <30s. • Theoretical resolution: 0.01%. • On-chip proprietary temperature compensation algorithms, enabling better data quality in wide temperature detection, and HAZMAT response. and humidity ranges. • Warm-up time from a cold start: about 45s. • Zero drift: <+0.05% VOL/month. • Estimated service life: >5 years. • Operating temperature: -20 to 50°C. • Operating humidity: 0~95%RH. • The default target gas is methane (CH4). To measure other types of hydrocarbon, users need to adjust the sensitivity correction factor of the module. | For general environmental monitoring, oil and gas leak detection, and HAZMAT response. |
Wide-range CO2 Sensing Module | • Detection method: Non-dispersive infrared (NDIR). • Range: 0~5%VOL/50000ppm. • Detection limit: 0.01%/100ppm. • Accuracy: ±10%FS. • Repeatability: <2%FS. • Response time (t90): <30s. • Theoretical resolution: 0.01%/100ppm. • On-chip proprietary temperature compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Warm-up time from a cold start: about 45s. • Zero drift: <±0.05% VOL/month. • Estimated service life: >5 years. • Operating temperature: -20 to 50°C. • Operating humidity: 0~95%RH. | For HAZMAT response. |
Wide-range Volatile Organic Compounds (TVOC) Sensing Module | • Detection method: photoionisation detection (PID). • Target gases: Total volatile organic compounds (TVOC) with ionisation potential energies <10.6eV. • Range: 0~50ppm (isobutylene). • Detection limit: 5ppb. • Repeatability: <4%FS. • Response time (t90): <3s (diffusion mode). • On-chip proprietary, enabling better data quality in wide temperature and humidity ranges. • Theoretical resolution: ~1ppb. • Warm-up time from a cold start: ~5minutes. • Estimated service life: 5000 working hours. • Operating temperature: -40 to 55°C. • Operating humidity: 0-95%RH. • Humidity has almost no effect on the measurement in 0~75%RH; • The default target gas is isobutylene. To measure other types of VOC, users need to adjust the sensitivity correction factor of the module. | For general environmental monitoring, oil and gas leak detection, and HAZMAT response. |
HCHO Sensing Module | • Detection method: Solid-state electrochemistry. • Range: 0~100ppm. • Detection limit: 1ppm. • Accuracy: ±5%FS. • Repeatability: <2%. • Overall response time (t90): <3s (0 to 50ppm); • Theoretical resolution: ~0.1ppm; • On-chip proprietary temperature compensation algorithms, enabling better data quality in wide temperature and humidity ranges. • Est. service life: >36months. • Warm-up time from a cold start: ≈5minutes. • Operating temperature: -40 to 55°C (Note that the module may require readjustment on its zero point due to changes in operating temperature or working environment). • Operating humidity: 15-95%RH (non-condensing). |